TIENet 310 Ultrasonic Level Sensor with Signature Flow Meter

  • MCERTS MCERTS certified meeting the Environment Agency's performance requirements
  • ATEX ATEX Group II, Category 1G (zone 0), T4
  • Non-contact liquid level measurement Built-in level-to-flow conversions
Product Options

Overview

TIENet 310 Ex Ultrasonic Level Sensor

The Signature® Flow Meter uses the TIENet 310 Ex Device to provide non-contact liquid level measurement. The flow meter has built-in level-to-flow conversions that cover the majority of open channel flow measurement situations. The ultrasonic sensor is mounted over the flow stream. The flow meter measures the time interval between transmission of a sound pulse from the sensor, and receiving its echo off the surface of the liquid, to determine the level of the stream.


Signture Flow Meter

The new Isco Signature flow meter is designed for open channel flow monitoring applications. It supports flow measurement technologies including ultrasonic and bubbler. Flow can be calculated using standard open channel level-to-flow conversions, level to flow data points as well as user-defined equations, depending on the needs of the application.
Unique features provided by the Signature flow meter help to verify data integrity. It logs key events such as power outages and changes in calibration to validate data accuracy. Stored data can be easily reviewed to detect any type of data alteration.

Manufactured by: Teledyne ISCO
Model: TIENet 310 Ultrasonic Level Sensor with Signature Flow Meter

Services offered for this product

  • Sales
  • Rental
  • Installation
  • Calibration
  • Maintenance

Specification

310 Ex TIENet Device Specifications a
Sensor Dimensions 3.63" ø x 4" tall (9.1cm ø x 10.2cm tall)
Cable Length 10 or 23 meters standard (32.8 or 75.5ft) standard
Mounting Attachment 3 /4" NPT Pipe thread nipple w/ Conduit lock nut
Weight 4 lbs (1.8 kg)
Body Material PVDF
Enclosure g IP68 when connected and properly sealed with cord-grip fittin

Temperature Range

Operating (compensated)

Storage

Hazardous Locations

-22 to 140°F (-30 to 60°C)

-40 to 158°F (-40 to 70°C)

-40 to 140°F (-40 to 60°C)

Measurement Range

Minimum: 1 ft (0.3m) from sensor to liquid surface

Maximum: 11 ft (3.3m) from sensor to liquid at minimum level

Measurement Accuracy

at 72 °F (22 °C)

±0.02ft (0.006m) at 1ft level change or less;

±0.03ft (0.009m) at greater than 1ft level change

Temperature Coefficient within compensated range

± 0.0002 x D (m) per degree C

± 0.00011 x D (ft) per degree F

(Where D is the distance from the transducer to the liquid surface)

Beam Angle 10° (5° From center line)
Ultrasonic Signal 50KHz
Certifications

Group II, Category 1G (zone 0), T4, [ATEX and IECEx]

Class I, Division 1 (and Zone 0), T4