Aurora RM - Rack Mount Moisture Analyser

  • Suitable for Installation in Safe Areas
Product Options

Overview

The Panametrics Aurora RM is a rack mount moisture analyser which can be used to measure moisture in a range of background gases. The Aurora RM uses tunable diode laser absorption spectroscopy (TDLAS) for accurate moisture measurement in a variety of environmental conditions.

Manufactured by: Panametrics
Model: Aurora RM - Rack Mount Moisture Analyser

Services offered for this product

  • Sales
  • Rental
  • Installation
  • Calibration
  • Maintenance

Specification

Range
Range 0 to 5000 ppmv
For CO2 applications: 0 to 1000 ppmv
Lower Detection Point 2 ppm
For CO2 applications: 20 ppm (-55.3°C)
Dew/Frost Point -97°F to 27.3°F (-71°C to -2.6°C) frost point @ STP of 25°C, 14.696 psia
Process Dew/Frost Point Process or equivalent dew point/ frost point by calculation with process pressure signal (4-20 mA) or constant
Absolute Humidity 0.095 to 237 lbs/MMSFC
(1.52 to 3,803 mg/m3)
Accuracy
Moisture reading (parts per million by volume)

±1% of reading or ±2 ppmv, whichever is greater; for > 1000 ppmv, ±5% of reading

For CO2 applications:
±3% of reading or ±5 ppmv

For H2 recycle applications: ±1% of reading or ±2 ppmv (for up to ±5% H2 and ±1% C2H6 variation from nominal calibration composition)

(Background conditions for individual instrument calibrated accuracy provided in Certificate of Conformance. Accuracy of other parameters derived from ppmv.)

Repeatability ±0.2 ppmv or ±0.1%, whichever is greater; For CO2 applications: ±1.0 ppmv or ±0.5%, whichever is greater
Calibration Certification NIST or equivalent NMI traceable certification
Calibration Options Nitrogen, standard natural gas and 3 customizable calibration curves
Response Time
Response Time Optical system <2 seconds
System Response The system response is dependent on the length of sample tubing, sample system components, flow rate and pressure, as well as the change in moisture concentration.
Pressure
Operating Sample Cell Pressure 69 to 172 Kpa (10 to 25 psia)
Process Pressure 102 psig (0.69MPa)
Higher pressure available with application of additional sampling system components
Flow Rate
Sample Cell Flow Rate 10 to 60 SLH (0.4 to 2 SCFH );
30 SLH (1 SCFH) nominal
By-pass Fast Loop 5 to 10X of flowrate through sample cell, available upon request
I/O
Display Backlit transflective LC display. Three programable simultaneous parameters. Alphanumeric status and diagnostic display. LEDs for power, laser temperature stability, keypad lockout
Power analyser: 100-240 VAC, 50-60 Hz
Analog Outputs Three 0/4-20 mA DC (source) with 500 ohm load. User programmable for any parameter and scalable. Complies with NAMUR protocol for analog signals
Analog Input Loop powered 4-20 mA input for remote pressure transmitter. Aurora supplies 24 VDC.
Digital Interfaces Two programmable digital communications ports RS232, RS485 with multidrop capability and assignable address, MODBUS RTU protocol. One Ethernet (RJ-45) port: Modbus TCP/IP protocol
User Interface Tactile keypad with status/fault indicator LEDs
Laser Class 1 product. Conforms to IEC 60825-1. Edition 2.0 Safety of Laser Products
Enclosure
Ingress Protection IP 20
Net Weight 19 kg (41 lbs)
Dimensions (H x L x W) 7” H (177 mm) X 19” W (483 mm) X 22” D (559 mm)
Temperature
Operating -20 to 50°C (-4 to 122°F)
Storage -20 to 70°C (-4 to 158°F)
Certification
USA/Canada General Purpose
(No hazardous area certification)
EU and Elsewhere General Purpose; Complies with 2014/30/EU EMC and 2014/35/EU LVD harmonizing directives

Downloads